S/5 M-PRESTN Modules
13
Document No. 8005571-1
Memory
NIBP program memory (processor flash memory) size is 512k
×
8. The processor has 4 kBytes RAM
and there is also an external RAM memory the size of which is 128k x 8. Variable values of the NIBP
measurement are stored into the external RAM. The EEPROM size is 512 x 8 and it is used to store
the calibration values for the pressure transducers, the pulse valve constants gained during
measurements, the PC board identification, and module serial number.
Software control
Software controls valves and pump. In addition to the individual on/off signals for each component
there is a common power switch for the valves and the pump that can be used at pump/valve
failures.
In addition to external RS485 reset line the microprocessor system is equipped with its own power-
up reset. See the section in the ECG board’s description:
Safety circuit
The NIBP board is equipped with an independent safety circuit to disconnect supply voltages from
the pump and the valves if the cuff has been pressurized longer than the preset maximum
measurement time, or if the pressure of the cuff is inflated over the specified pressure limit. The
maximum measurement time values and pressure limits for different measurement modes have
been specified in the technical specification section of this manual.
Pneumatics
Pneumatics of PRESTN module has the following parts:
•
Intake air filter
; for preventing dust and other parts to enter the air pump and the
valves.
•
Air pump
: for pumping the measuring pressure of the cuff.
•
(
Pulse) Valve
; for producing a linear pressure fall (bleeding) in order to measure the
blood pressure of the patient.
Note that there has been used also two other names
Valve
and
Set valve
to
designate pulse valve in service menu.
•
Safety valve
; The safety valve has been intended to be used for deflating the cuff in
single fault case, i.e. to prevent too long measurement time or too high inflation
pressure of the cuff.
Note that there has been used also
Exh2 valve
to designate the
Safety valve
in
service menu.
•
Main pressure sensor
; for measuring the pressure of the blood pressure cuff and
the pressure fluctuations caused by arterial wall movement.
•
Safety pressure sensor
for detection of cuff hose type, cuff loose, cuff occlusion
situations etc.and recognising the pressure sensor fault.
•
Cuff connector
; for connecting the cuff.
Содержание S/5 PRESTN
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Страница 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Страница 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Страница 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Страница 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...