S/5 M-PRESTN Modules
41
Document No. 8005571-1
4. Pump the following pressures to manometer and check the difference between the
manometer and monitor pressure display (The zeroing offset is automatically subtracted from
the pressure readings).
Table 4
NIBP calibration check pressures
Pressure Max.
error
Example
0 mmHg
±
5 mmHg (=zero offset)
-1
100 mmHg
100
±
2 mmHg
100
±
2
200 mmHg
200
±
3 mmHg
200
±
3
If the error of pressure channel B1 is larger than specified above, the module should be
recalibrated. The error of B2 is allowed to be even twice as large because it has no effect on blood
pressure measurement accuracy. However, we recommend recalibrating the module when the error
of B2 is larger than specified above to ensure best possible operation.
Calibration
1. Enter
Calibration
menu.
2.
Remove hoses from front panel connector to enable proper zeroing.
3. Select
Calibration
. If it is not available, perform the steps A, B, and C.
NOTE: Do not pull out the NIBP module from the monitor frame. The module must be in the frame
during the whole procedure.
A. Press NIBP module buttons
Auto ON/OFF
and
Start Cancel
simultaneously for 3 seconds to
enable the calibration. This enables menu selection
Protection
. The message ‘Calibration
switch ON!’ appears.
B. Select
Protection OFF
in the Calibration menu and press the ComWheel.
C. Press the buttons again for 3 seconds. Menu selection
Calibration
is now enabled, and
Protection
is disabled. When the calibration is enabled, a message ‘Calibration not
protected’ appears.
•
Start Calibration by pressing the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in
the NIBP message field. After this a pressure bar and text ‘Calibrating’ will appear.
•
Connect an external mercury manometer with pump to module through the both tubes of the
hose - both transducers B1 and B2 must be calibrated simultaneously. Pump up to a
pressure about 200 mmHg according to the manometer. Calibration is possible in the range
150 to 250 mmHg.
•
Verify that both pressure values in the prompt field match the manometer reading. If not,
adjust by turning the ComWheel. When the values of the pressure bar and the manometer
are equal, press the ComWheel to confirm the calibration. The message ‘Calibrated’ will
appear onto the NIBP digit field after a few seconds, which means that the calibration
succeeded, and the new calibration data is saved in EEPROM.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the
ComWheel, even in cases where the value would be correct. For example change the value one step
higher and then back one step lower. “Calibrated” text should appear in the display. This ensures
that the calibration procedure is correctly registered and stored by the module.
Содержание S/5 PRESTN
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Страница 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Страница 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Страница 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Страница 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...