![Datex-Ohmeda S/5 PRESTN Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/datex-ohmeda/s-5-prestn/s-5-prestn_technical-reference-manual-slot_513816017.webp)
S/5 M-PRESTN Modules
11
Document No. 8005571-1
2.2 Main components
2.2.1 M-PRESTN/-RESTN/-PRETN modules
NIBP
ECG
SpO
T1
T2
P1
P2
2
Auto
On/Off
Start
Cancel
Zero
P1
Zero
P2
Figure 3
Front panel of M-PRESTN
The
M-PRESTN, M-RESTN, and M-PRETN
modules
contain three main PC boards, the STP board,
the ECG board, and the NIBP board. They work independently. Each of these has their own
processor and software in the processor flash memory.
There are two small boards, the SP input and the ECG input board attached to the front panel of the
module. The front panel has seven connectors and four keys. The connectors are two for
temperature measurement, two for invasive blood pressure measurement, one for ECG, one for
NIBP, and one for SpO
2
measurement. The keys are for NIBP Auto On/Off, NIBP Start/Cancel, P1
zero, and P2 zero.