S/5 M-PRESTN Modules
35
Document No. 8005571-1
Restore the normal monitoring screen by pressing the key
Normal Screen
.
Connect cables from the channels BP1 and BP2 to the module connectors P1 and P2. Zero
the InvBP channels by pressing the keys ZERO P1 and ZERO P2 on the module front panel.
BP - 1 - WAVE - ART
2 - WAVE - CVP
Check that appropriate InvBP waveforms are shown and the InvBP values are approximately
120/80 (
±
3 mmHg) for the channel P1 and 15/10 (
±
2 mmHg) for the channel P2.
Check that HR- value is calculated from P1 when ECG is not measured (ECG cable
disconnected).
"
SpO
2
measurement
23. Check that the message ‘No probe’ is shown when no SpO
2
sensor is connected to the
module. Connect a SpO
2
finger probe to the module. Check that the message ‘Probe off’ is
shown when the probe is not connected to a finger.
"
24.
Connect the SpO
2
probe onto your finger. Check that the reading of
95-99 and SpO
2
waveform appears. Check that HR- value is calculated from SpO
2
when
ECG and InvBP (P1) are not measured.
"
Non Invasive Blood Pressure measurement
25. Enter the NIBP module service menu:
Parameters - NIBP
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 5 per second. Check also that the NIBP board memories have passed the
internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ show all OK.
"
26. Check the front panel membrane keys.
Select
Buttons/Leds
.
Press each of the two NIBP related membrane keys at least for one second. Check that the
pressed key is identified, i.e. the corresponding text changes from OFF to ON in the menu
when the key is released back up again.
"
27.
Check the pump and valves.
Highlight
Pneumatics
from the NIBP menu. Connect a pressure manometer to the NIBP
Содержание S/5 PRESTN
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Страница 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Страница 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Страница 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Страница 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...