9.
Connect a 12-lead ECG trunk cable without a lead set to the module. Check that the
message "Leads off" is displayed on the screen.
"
10.
Connect both 5-leadwire sets to the trunk cable. Connect limb lead electrodes and one
electrode from the chest lead set to the same potential. Check that the ‘Cable type’ shows
10 lead.
"
11.
Change the 3-leadwire set to the trunk cable. Check that all the electrodes show OFF in the
service menu and the message ‘Leads Off’ is shown on the screen.
Connect all the leads together, for example to a suitable screwdriver. Check that all the
electrodes show ON and the message ‘Asystole’ appears.
Disconnect one of the leads and check that the corresponding electrode in the service menu
shows OFF within 10 seconds from the disconnection, and then reconnect the lead. Check
the rest of the leads using the same method.
NOTE: When the ground lead (black) is disconnected all the electrodes should show OFF.
NOTE: The ‘Asystole’ and ‘Different leads off’ messages are shown using certain priority, so
even when one of the leads is disconnected, the lead related ‘Leads off’ message may not
appear onto the screen.
NOTE: When RA, LA or LL electrode is disconnected, all six V electrodes show OFF.
NOTE: With PRESTN/RESTN/PRETN modules and 5 lead cable the state of V2, V3, V4, V5
and V6 electrodes follow the state of the V electrode.
Connect the leads to a patient simulator.
Perform the settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
ECG - BASE - BPM - 160
PACE - WAVE - NSR
Check that normal ECG waveform is shown, the HR -value is 160 (±5) and the ‘Pacer count’ -
value is not increasing in the service menu.
ECG - PACE - WAVE - ASNC
Check that pacemaker spikes are shown on the ECG waveform, the ‘HR’ -value changes to
75 (±5) and the ‘Pacer count’ -value is increasing according to shown pacemaker spikes.
Set the pacemaker option off:
ECG - PACE - WAVE - NSR
"
Содержание S/5 PRESTN
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Страница 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Страница 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Страница 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Страница 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...