APPENDIX B, EMC tables, S/5 Compact Monitors
Table 4 Recommended separation distances between portable and mobile RF
communications equipment and the S/5™ CM (1
Recommended separation distances between portable and mobile RF communications
equipment and the S/5™ CM (1 .
The S/5™ CM (1 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the S/5™ CM (1 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the S/5™ CM (1 as recommended below, according to the maximum output power of the communications
equipment.
Separation distance according to frequency of transmitter
m
Rated maximum
output power of
transmitter
W
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
d = 3.5 P
(2
800 MHz to 2.5 GHz
d = 2.3 P
d = 7.0 P
(2
0.01 0.12
0.12
0.35
(2
0.23
0.70
(2
0.1 0.38
0.38
1.1
(2
0.73
2.2
(2
1 1.2
1.2
3.5
(2
2.3
7.0
(2
10 3.8
3.8
11
(2
7.3
22
(2
100 12
12
35
(2
23
70
(2
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(1
S/5™ CM equipped with M-NESTPR or M-NE12STPR, M-ENTROPY, M-BIS, M-MINIC and N-DIS.
(2
For M-BIS.
Document No. M1031508
B-5(6)
Содержание S/5 Compact Anesthesia Monitor
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Страница 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Страница 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Страница 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Страница 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Страница 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...