General Service Guide
2. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement.
Check the module identifies the cuff, i.e. the text ‘Adult ’ appears on the NIBP digit field for a
short time.
Check that the module gives a reasonable measurement result.
"
5.3.18 Nellcor Compatible Saturation module, M-NSAT
1. Connect a Nellcor SpO2 finger sensor to the module. Check that the message ‘Pulse search ’
is shown. Check also that the shown message changes to ‘Check probe ’ within 30 seconds.
2. Attach the SpO2 sensor on your finger. Check that a reading of 95-100 and a proper SpO2
waveform appear.
"
5.3.19 Datex-Ohmeda Oxygen Saturation module, M-OSAT
1.
Connect a Datex-Ohmeda OxyLead Cable compatible sensor ot Integrated sensor to the
module. Check that the message ‘Pulse search ’ is shown. Check also that the shown
message changes to ‘Check probe ’ within 30 seconds.
2.
Attach the SpO2 sensor on your finger. Check that a reading of 95-100 and a proper SpO2
waveform appear.
"
5.3.20 EEG Module, M-EEG
1. Check that the EEG module is recognized by disconnecting it shortly.
"
5.3.21 BIS-Module, M-BIS
1.
Connect the BIS module to the monitor frame.
Check that 'Cable off' is displayed in the BIS waveform field.
2.
Connect the DSC to the module.
Check that 'No sensor' message appears to the waveform field.
3.
Check the functioning of the front panel membrane keys.
Pressing
BIS
key should open BIS Setup page. Check that 'Test DSC' shows PASS.
"
NOTE If sensor is not connected, the Check Sensor key is inoperative.
Document No. M1031508
47
Содержание S/5 Compact Anesthesia Monitor
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Страница 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Страница 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Страница 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Страница 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Страница 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...