General Service Guide
Item Settings
Page counter
Actual value displayed
Ritech Medium
Standby Enable
4.5.5 Connection to Epson EPL-5500 printer
By default the printer has only a parallel port, but it is possible to add a serial module with a serial
port for the connection. The printer and the serial module (C823341) should be purchased from a
local Epson dealer. Also the printer cable should be purchased from a local Epson dealer.
WARNING
Connecting the power supply cord of the printer to the wall socket may cause
the printer leakage current to exceed the limit specified for medical equipment.
Always connect the printer to an appropriate separating transformer.
Printer setup
NOTE: For a detailed description, please refer to the documentation provided with the printer.
Serial connection
1.
To select the serial mode open the printer cover, remove the side cover, and the shield
cover. Check that the jumper switch and the side switches on the serial module are in the
serial interface position.
Parallel connection and serial connection
2. Insert the photoconductor unit, and the developer cartridge into the printer.
3. Plug the printer into mains supply, and connect the printer cable.
Serial connection
4. Hold down the button on the control panel, and turn on the printer. Continue to hold down the
button until both lights on the control panel turn off, then release the button. Make sure that
the green light is flashing, indicating that the printer is in the default setting mode. Input the
following code numbers:
Code Option
124
Set baud rate to 19200
111
Exit the default setting mode
e.g. for 124 push the button on the control panel first once, then twice, and then four times.
Parallel Connection and Serial Connection
Load paper into the paper tray and open the top paper support.
4.6 Interfacing computer
A computer is connected to the serial port connector X11 on the frame unit.
Contact your authorized Datex-Ohmeda distributor for further advice on computer interface.
Document No. M1031508
37
Содержание S/5 Compact Anesthesia Monitor
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Страница 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Страница 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Страница 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Страница 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Страница 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...