Datex-Ohmeda S/5 Compact Anesthesia and Critical Care Monitor
Cautions
CAUTION
A CAUTION indicates a condition that may lead to equipment damage or
malfunction.
Installation
•
Leave space around the monitor to allow proper ventilation.
•
Ensure that the module is properly orientated (i.e. module release latch facing downward)
before insertion.
Before use
•
Allow two minutes for warm-up and note any error messages or deviations from normal
operation.
•
Clean the fan dust filters once a month or whenever necessary.
•
Do not connect a sampling line to the female Patient Spirometry connector while the other
end of the sampling line is connected to the D-fend water trap. The pressure in the gas
sampling system may cause damage to the PVX unit pressure transducer.
Autoclaving and sterilizing
•
Do not autoclave any part of the monitor.
•
Do not gas sterilize the modules.
Cleaning and service
•
Do not use ammonia, phenol, or acetone based cleaners. These cleaners may damage the
monitor surface.
•
Do not immerse the monitor in any liquid. Do not allow liquid to enter the monitor or
modules.
•
Electrostatic discharge through the PC boards may damage the components. Before
handling PC boards, wear a static control wrist strap. Handle all PC boards by their non-
conductive edges and use anti-static containers when transporting them.
•
Do not break or bypass the patient isolation barrier when testing PC boards.
Special components
•
Special components are used in these monitors that are vital to assure reliability and safety.
Datex-Ohmeda assumes no responsibility for damage, if replacement components not
approved by Datex-Ohmeda are used.
•
A lithium battery on the Central Processing Board. Dispose of the faulty IC containing the
battery according to local regulations.
Document No. M1031508
8
Содержание S/5 Compact Anesthesia Monitor
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Страница 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Страница 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Страница 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Страница 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Страница 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...