Chapter 3 – Printer Operation
19
3.7.3 Standard Calibration
The Standard Calibration can be performed using the DMXConfig Utility,
(see section 3.5 for more information
on DMXConfig)
or using the front panel buttons
(see Appendix B for more information).
Once you have installed the DMXConfig utility and the printer is properly loaded with media:
1. Connect the host to the printer with a serial
or parallel cable.
2. Turn on the printer.
3. Launch the DMXConfig utility.
4. Query the printer by using the ‘Query Printer’
toolbar button (top-left). This will connect to
the printer and get the current printer
settings.
5. Select the “Sensor Calibration” Tab and then
click the “Media Calibration Wizard” button.
When prompted click “OK” to start the
calibration wizard.
6. The Calibration Wizard will now prompt you to
‘Load Stock’. Be sure the media is properly
loaded in the printer. Close the printhead and
click “OK”.