Ulyxe PL™
52
6
S
TE
C
OUT
PU
FA
ULT
The Peltier module voltage software
level was exceeded. The value is set
at 10V and it is displayed in the
TEC
Voltage
field in the SYSTEM DATA
section. The system turns off the
entire power section (laser diode,
Peltier module, RF driver, etc.) since
the internal card is malfunctioning.
Turn off, wait 30/60 minutes and
restart. Read the value displayed in
the
TEC Voltage
field in the
SYSTEM MONITOR section. In
case the problem reoccurs, carry
out the data log with the service
tool and contact technical
assistance.
S
PWM
CAR
TE
MP
ER
FA
ULT
The temperature software level on
the power section of the internal
card was exceeded. The value is set
at 65℃ . The system turns off the
entire power section (laser diode,
Peltier module, RF driver, etc.) since
the internal heat cannot be
dissipated. The problem may be
caused by critical environmental
conditions (high outside
temperature, insufficient air flowing
into the fans, etc.) or the internal
dissipation is not efficient (poor
thermal contact between the internal
devices).
Turn off, wait 30/60 minutes and
restart.
Make sure the environmental
conditions comply with
specifications. Read the value
displayed in the
Card Temp.
field in
the SYSTEM MONITOR section,
checking to see if it quickly reaches
65°C. In case the problem re-
occurs, carry out the data log with
the service tool and contact
technical assistance.
S
HEAT
SIN
K
HIG
H TE
M
FA
ULT
The temperature software level on
the central heat sink was exceeded.
The value is set at 60°C. The
system turns off the entire power
section (laser diode, Peltier module,
RF driver, etc.) since the internal
heat cannot be dissipated. The
problem may be caused by critical
environmental conditions (high
outside temperature, insufficient air
flowing into the fans, etc.) or by a
malfunction of the cooling fans.
Turn off, wait 30/60 minutes and
restart. Make sure the
environmental conditions comply
with specifications. Read the value
displayed in the Heat sink T. field in
the SYSTEM MONITOR section,
checking to see if it quickly reaches
60°C. In case the problem re-
occurs, carry out the data log with
the service tool and contact
technical assistance.
S
SYSTEM
HIG
H TE
M
FA
ULT
The environment temperature
software level was exceeded. The
value is set at 40°C. The system
turns off the entire power section
(laser diode, Peltier module, RF
driver, etc.) since the internal heat
cannot be dissipated.
Turn off Ulyxe PL
™
and make sure
the environment temperature
complies with the specifications
before you turn it back on. Read the
value displayed in the
System
Temp.
field in the SYSTEM
MONITOR section, checking to see
if it is near the 40°C limit. In case
the problem re-occurs, carry out the
data log with the service tool and
contact technical assistance.
Содержание Ulyxe PL
Страница 1: ...USER S MANUAL Ulyxe PLTM ...
Страница 27: ...TECHNICAL SPECIFICATIONS 27 2 EXAMPLE OF CONNECTION TO PLC Figure 18 Example of connection to PLC ...
Страница 56: ...56 LABELS LOCATION Figure 44 External labels location ...
Страница 66: ...66 APPENDIX D MECHANICAL DRAWINGS NOTE Please refer to Datalogic website for detailed drawings ...
Страница 68: ...68 ...