DEVICE MAINTENANCE
116
SAFERY LASER SCANNER
Fast Replacement of a Slave device
The Fast Replacement of a Slave device can occur only when a damaged Slave device is
replaced with a new Slave device of the same model, and therefore only the serial num-
bers do not match.
In this case, the new device can be aligned with the Memory Group topology by
restor-
ing the configuration
from the Memory Group to the device. To restore the configura-
tion from the Memory Group:
1. select the Restore option. The Slave device display will show the “CHECK MASTER”
message. The Master device display will show the “RUN TEST MODE” message;
2. select “ENTER TEST” using the keypad of the Master device to start the test proce-
dure;
3. at test completion, select “EXIT TEST”;
4. the user is now prompted to
validate
or
reject
the configuration.
Alternatively, a new configuration via DLSentinel can be created (refer to the DLSentinel
User’s Manual).
NOTE
•
a new Slave device is added to the current cluster, OR
•
a damaged Slave device is replaced with a different Slave model (e.g.
a 3-meter model is replaced with a 5-meter model).
When the cluster size or models do not match, a new configuration via
DLSentinel must be created (refer to the DLSentinel User’s Manual). This is
the case when:
Содержание Laser Sentinel
Страница 1: ...Laser Sentinel INSTRUCTION MANUAL Safety Laser Scanner ...
Страница 18: ...GENERAL VIEW xviii SAFERY LASER SCANNER ...
Страница 140: ...122 SAFERY LASER SCANNER APPENDIX B OVERALL DIMENSIONS LASER SENTINEL STAND ALONE MODEL ...
Страница 141: ...LASER SENTINEL MASTER MODEL INSTRUCTION MANUAL 123 LASER SENTINEL MASTER MODEL LASER SENTINEL SLAVE MODEL ...
Страница 149: ......
Страница 150: ......
Страница 151: ......