Linea Lite GigE Series Camera
Camera Performance and Features
•
45
FPN Correction
In general, factory FPN correction is sufficient for most applications. However, if a new FPN
correction is required a user set FPN correction can be applied.
To perform FPN Correction
:
Step 1.
Cover the lens (place the sensor in dark).
Step 2.
to Off and check the line profile / histogram. If some, or
all, of the pixels outputs are zero, then you should adjust the
ensure that all pixel output is above zero.
Step 3.
to Calibration and select the required user set from
Field Correction Current Active Set
Step 4.
Flat Field Calibration Sample Size
feature to 4096 or 2048. The 4096 option
returns more accurate results whereas the 2048 option reduces calculating time. In general, the
2048 results are still satisfactory for most applications. The
Step 5.
In the Calibrate FPN feature, click the
Press…
command to perform the calibration.
Preliminary