VII. Abnormalities and Troubleshooting
72
Dakewe (Shenzhen) Medical Equipment Co., Ltd.
www.dakewemedical.com
Figure 7- 6 Taking out the slide
Figure 7- 7 Abnormal operation code: OX09
4. OX09 error
Abnormality Description
The transfer push rod stops running during operation. A prompt box pops up in the main interface (Figure
7-7).
Cause analysis
The input hopper is not properly inserted;
Too many slides in the input hopper;
The height difference of the two sides the piles of the slides is greater than 1.5 mm;
Slide sticking; Broken edge of slide;
Irregularly shaped slide;
Damaged paint surface of slide;
The slide is placed upside down;
Bevelled slides are used;
Debris between slides;
Debris on the print platform;
Debris on the input hopper holder.
Resolution
1. Reset the instrument: Check whether the input hopper is properly inserted; click “Reset” after it is
properly inserted.
2. If the resetting fails, take out the stuck slide:
(1) Open the maintenance door, and gently move the transfer push head in the opposite direction of the
stuck slide (Figure 7-8);
Note: Move the transfer push rod to the right when printing the slide on the right, and move to the left
when printing the slide on the left;
(2) Move the slide to the print position (Figure 7-9);