Error No.
Message
Possible cause
Corrective action
ERROR 3
STERILIZATION
FAIL
The temperature when
sterilizing the blade
dropped below the set
point.
•
Repeat the operation with a
new blade.
•
Call Service if the problem
persists.
ERROR 5
RETRACT
MOTOR FAIL
The blade retract operation
failed to retract the blade in
the specified time.
•
Check for obstructions.
•
Switch off the power to the
instrument.
•
Switch on the power to the
instrument and follow the
display messages.
•
If the motor fails to move,
contact Service.
ERROR 6
CUT MOTOR
FAIL
The cut operation failed to
reach the cut position in the
specified time.
•
Switch off the power to the
instrument.
•
Switch on the power to the
instrument and follow the
display messages.
•
If the motor fails to move,
contact Service.
ERROR 7
ALIGN MOTOR
FAIL
The align operation failed to
align the cut tubes in the
specified time.
•
Switch off the power to the
instrument.
•
Switch on the power to the
instrument and follow the
display messages.
•
If the motor fails to move,
contact Service.
ERROR 8
LIFTOUT
MOTOR FAIL
The blade liftout operation
failed to bring the blade to
the top position in the
specified time.
•
Switch off the power to the
instrument.
•
Switch on the power to the
instrument and follow the
display messages.
•
If the motor fails to move,
contact Service.
ERROR 9
TEMPERATURE
SENSOR FAIL
The infrared temperature
sensor has failed.
Contact Service to replace and
recalibrate the infrared tempera-
ture sensor.
7 Troubleshooting
Sterile Tube Fuser - Wet Operating Instructions 29024010 AD
67