NOTICE
This equipment is not intended for use in residential environments
and may not provide adequate protection to radio reception in
such environments.
NOTICE
Only use PTFE coated cutting blades (BLADES-IR/50) specified for
use with Sterile Tube Fuser - Wet. For optimal use, do not reuse
blades.
Personal protection
WARNING
Always use appropriate personal protective equipment during
installation, operation, inspection, maintenance and cleaning of
Sterile Tube Fuser - Wet.
WARNING
Spread of biological agents. The operator must take all neces-
sary actions to avoid spreading hazardous biological agents. The
facility must comply with the national code of practice for
biosafety.
WARNING
Do not attempt to weld two different tube materials. This may lead
to poor weld/no weld and spillage.
WARNING
Do not attempt to access moving parts and heating zone during
welding operation by any means.
2 Safety instructions
2.1 Safety precautions
12
Sterile Tube Fuser - Wet Operating Instructions 29024010 AD