When the process has fi nished and the Zero calibration was accepted, press ENTER to Exit and
remove the cylinder of zero gas. Refer to Section
7.3.3 Zero - Requires Override
if the Zero calibration
was not successful.
NOTE:
After 5 minutes of inactivity, the display will return to the normal operation.
Step 2
Set the span calibration gas level. It is important to make sure that the span gas level value matches
the calibration gas concentration you are using to calibrate the sensor.
The factory default calibration gas concentrations are:
SENSOR GAS TYPE
CALIBRATION GAS LEVEL
Carbon Monoxide (CO)
100 ppm
Nitrogen Dioxide (NO
2
)
5 ppm
Combustibles
20% LEL
In the Sensor Confi guration Menu (passcode 3032), press either ARROW key to scroll through the
menu and select the channel you want calibrate. Press ENTER and use the ARROW key to scroll to
Calibrate Span. Press ENTER and use the ARROW keys in increase or decrease the span gas value.