In the Sensor C
onfi guration Menu (passcode 3032) press either ARROW key to scroll through the
menu and select the gas channel for which you want to see the diagnostic information. Press ENTER
and then the UP ARROW key to see this diagnostics screen.
ADC
The current Analog to Digital Count number
Reading
The fi nal gas reading after all fi lters
N
The new sensitivity in ADC per unit of the Reading (ie. 26 ADC/PPM)
O
The original sensitivity in ADC per unit of the Reading (ie. 8 ADC/PPM)
Zero ADC
The ADC used at Zero calibration
Span ADC
The ADC used at Span calibration
Zero
The gas reading at Zero (in the same units as the Reading)
Span
The gas reading at Span (in the same units as the Reading)