(see Section
8.4 Calibration Kit
and can also be ordered separately from CETCI under part number
CET-MW
.
To initiate non-intrusive calibration, touch the magnet to one of the marks on the side of the
enclosure and follow the screen prompts along with the instructions in this section.
NOTE:
The programming buttons inside the enclosure can be used instead of the magnetic wand,
if preferred.
7.2 Calibrating the Internal Sensor(s)
NOTE:
If calibrating an Oxygen sensor, refer to Section
7.5 Calibrating an Oxygen Sensor
.
To calibrate the internal sensor(s), the user must go through the following steps:
Step 1
In the Sensor C
onfi guration Menu (passcode 3032), press either ARROW key to scroll through the
menu and select the channel you want calibrate. Press ENTER and use the ARROW key to scroll to
Calibrate Zero. Press ENTER and use the ARROW keys to enter a value of 0.
Attach the regulator to the cylinder of zero air, insert calibration adapter into the sensor opening on
the front of the enclosure door (see Section
7.3.1 Calibration Adapter Plug
if the adapter will not fi t).
Press Enter to start the Zero calibration. The display will show a countdown from 18 seconds.