LPT-M - Operation Manual
NOTE:
The LOW, MID and HIGH setpoints of the alarm levels are confi gured in the Sensor
Confi guration Menu. Refer to Section
6.11 Setting Channel Alarm Setpoints
.
In the Hardware Confi guration Menu (passcode 1001), press either ARROW to scroll through the
menu and select Program Relay. Press ENTER and use the ARROW keys to scroll through the menu
and select Relay Alarms. Press ENTER and use the ARROW keys to scroll through the choices. Press
Enter to save and Exit.
6.10 Test Functions
6.10.1 Test Audible (Buzzer)
NOTE:
Before testing the audible alarm/sounding the buzzer, warn people in the vicinity of the
sound so unnecessary distress or response is not caused.
In the Hardware Confi guration Menu (passcode 1001), press either ARROW to scroll through the
menu and select Program Audible. Press ENTER and use the ARROW keys to scroll through the menu
and select Audible Test. Press ENTER and use the ARROW keys to increase or decrease the number of
seconds. You can enter a maximum number of 9999 seconds (2.7 hours).