When you wish to convert your RH-‐N95, call CPAC Equipment to
receive this emailed and downloadable software.
6.3.5.3 Settings Configuration
To reconfigure several internal settings, press “MENU” from “READY”
or “HEATER IS OFF” screen to display “SELECT ITEM, then press
“CONFIGURATION.” A scrolling menu is provided. A scroll bar on far
right of screen will allow advancement through the entirety of the
screen as depicted:
and
Six formats or operational parameters are presented of which the
operator may make changes. Note that none of these operator
configurations will affect treatment times and temperatures required
of the decontamination process. Press the desired category and follow
instructions on the screen for attribute selection. Once the option is
selected, the screen reverts to the “CONFIGURATION” listing screen. If
no more changes are to be made, press “DONE” to return to “READY”
screen.
6.3.5.4 Error Messages
In the event of a malfunction that may lead to the failure of meeting
decontamination parameters, error messages are displayed as
represented by the following red screen for those operations
indicating out of conformance with operating specifications. The
display of an error massage may cause the cycle to be terminated
prematurely. If so, the cycle will be documented as a failure and the
masks will need to be reprocessed.
The following error messages are to assist the operator in diagnosing
and correcting any problem that may arise:
Blower temperature sensor failed
Air temperature sensor failed
Return air temperature sensor failed
Chamber too hot
Return air too hot