12
RH-‐95 Unit
Nylon Lifting Straps
Cushioning
5.2
Location
5.2.1
The RH-‐N95 is designed for operation indoors in a protected,
relatively dust-‐free ambient temperature environment at a relative
humidity of < 80% up to 88
0
F (31
0
C).
5.2.2
Surface Support – Level and sufficient construction for weight of
loaded unit (approximately 100 pounds) and ancillary equipment.
5.3
Dimensions and Required Clearances
5.3.1
Outer dimension: 22.5” (572mm)W x 25” (625mm)D x 19.5” (492mm)H
5.3.2
Back of Unit to Back Wall: 3” (75mm)
5.3.3
Front Support Surface to Front Sterilizer: 1” (25mm)
5.3.4
Sides of Unit to Side Wall: 3” (75mm) –
See Note below
5.3.5
Distance Above the Unit: 3” (75mm)
NOTE: The main power on/off circuit breaker is located at the right rear of
the Rh-‐N95. Install the unit with enough clearance to permit reaching past
the right side of the unit to operate the switch. This is an infrequent need
since there is a front-‐panel On/Standby button, but the rear panel circuit
breaker should remain accessible. About 3.5-‐4.0 inches of side space is
sufficient reach-‐past for most individuals.
The cooling air supply filter (pictured on page 10”; rear elevation) requires
occasional rinsing to remove accumulated dust. Installation should include
consideration of access to that filter.