Error Codes and Suggested Resolutions
Service Manual
B-3
005
Category:
A/D
Alarm Priority:
Low
Description:
Temperature > 48°C for at least 30 consecutive seconds.
Resolution:
Replace the power PCBA. See
006
Category:
A/D
Alarm Priority:
Medium
Description:
Thermal cut-out switch has opened.
Resolution:
Check for:
•
Thermistor dislodged from hose duct adapter - See
.
If the thermistor is in position, replace the power PCBA. See
Note:
This error is also generated during the thermostat test as part of perfor-
mance verification.
011
Category:
A/D
Alarm Priority:
Medium
Description:
Thermistor temperature too low to measure.
Resolution:
Replace the thermistor. See
012
Category:
A/D
Alarm Priority:
Medium
Description:
Thermistor temperature too high to measure.
Resolution:
Replace the thermistor. See
Содержание WarmTouch
Страница 1: ...Service Manual WarmTouch Convective Warming Unit...
Страница 8: ...Page Left Intentionally Blank vi Service Manual...
Страница 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Страница 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Страница 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Страница 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Страница 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Страница 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Страница 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Страница 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Страница 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Страница 176: ...Page Left Intentionally Blank 10 58 Service Manual Internal Component Replacement Inside Front Enclosure...
Страница 232: ...Page Left Intentionally Blank 11 56 Service Manual Internal Component Replacement Inside Rear Enclosure...
Страница 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Страница 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Страница 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Страница 313: ......