Principles of Operation
Service Manual
6-7
Thermostat
The thermostat is located in the hose duct adapter near the base of the hose
(see
on page 11-10). It is connected to the power PCBA. The ther-
mostat provides thermal cut-out protection by opening at 49°C to 55°C (120°F
to 131°F), resulting in heater and fan shutdown.
Thermistor
The thermistor is located in the hose duct adapter just below the screen at the
base of the hose (see
page 11-7). It is connected to the power PCBA.
During operation, the temperature measured by the thermistor is monitored
by the A/D processor and relayed to the UI microprocessor for display on the
LCD. If the temperature does not remain within specification, the software
shuts down the heater and fan and generates an alarm.
Содержание WarmTouch
Страница 1: ...Service Manual WarmTouch Convective Warming Unit...
Страница 8: ...Page Left Intentionally Blank vi Service Manual...
Страница 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Страница 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Страница 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Страница 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Страница 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Страница 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Страница 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Страница 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Страница 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Страница 176: ...Page Left Intentionally Blank 10 58 Service Manual Internal Component Replacement Inside Front Enclosure...
Страница 232: ...Page Left Intentionally Blank 11 56 Service Manual Internal Component Replacement Inside Rear Enclosure...
Страница 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Страница 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Страница 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Страница 313: ......