Error Codes
B-14
Service Manual
184
Category:
UI
Alarm Priority:
Low
Description:
RTC NVRAM contained an illegal setpoint temperature.
Resolution:
Replace the clock battery. See
.
If this does not resolve the issue, replace the UI PCBA. See
.
186
Category:
UI
Alarm Priority:
Low
Description:
Error communicating with AD microprocessor across SPI bus.
Resolution:
Replace the ribbon cable. See
.
202
Category:
UI
Alarm Priority:
Low
Description:
UI microprocessor is receiving too many interrupts from the AD microprocessor.
Resolution:
This error may appear due to electromagnetic interference from other devices at
the customer site. Advise the customer to try increasing the distance between
devices, repositioning the cabling, or plugging the devices into separate outlet
circuit branches. See
Electromagnetic Compatibility (EMC)
on page 13-4 for addi-
tional guidance.
204
Category:
UI
Alarm Priority:
Low
Description:
Spurious interrupt from AD microprocessor.
Resolution:
This error may appear due to electromagnetic interference from other devices at
the customer site. Advise the customer to try increasing the distance between
devices, repositioning the cabling, or plugging the devices into separate outlet
circuit branches. See
Electromagnetic Compatibility (EMC)
on page 13-4 for addi-
tional guidance.
Содержание WarmTouch
Страница 1: ...Service Manual WarmTouch Convective Warming Unit...
Страница 8: ...Page Left Intentionally Blank vi Service Manual...
Страница 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Страница 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Страница 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Страница 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Страница 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Страница 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Страница 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Страница 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Страница 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Страница 176: ...Page Left Intentionally Blank 10 58 Service Manual Internal Component Replacement Inside Front Enclosure...
Страница 232: ...Page Left Intentionally Blank 11 56 Service Manual Internal Component Replacement Inside Rear Enclosure...
Страница 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Страница 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Страница 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Страница 313: ......