Alarms and Troubleshooting
3-14
Puritan Bennett 540™ Ventilator User’s Manual
3.8
Troubleshooting
3.8.1 Alarms
offers a guide to the most likely ventilator alarms, possible reasons for the alarms, and
corrective actions.
Caution
•
This manual tells you how to respond to ventilator alarms, but it does NOT tell you how to respond to the
patient.
•
To ensure proper servicing of the ventilator and reduce the possibility of physical injury, only qualified
personnel should attempt to service the Puritan Bennett 540™ Ventilator.
WARNING
When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem,
examine the patient first before examining the ventilator.
WARNING
Except for replacing the internal battery and cleaning the device, do not try to repair or otherwise service
the ventilator yourself, or modify the ventilator, its components, or accessories. Doing so might endanger
the patient, cause damage to the ventilator, and/or void your warranty. Only qualified service personnel
should attempt repair of the ventilator.
Table 3-2.
Alarms and Corrective Actions
Alarm Message or
Symptom
Possible Reason(s) For The Alarm Event
Corrective Action(s)
AC POWER
DISCONNECTION
AC (“mains”) power source cut off.
Cancel the alarm then check the supply
cable and/or the effective availability of a
voltage on the AC power (“mains”) port.
Starting with 12 – 30 VDC external power
supply.
Cancel the alarm.
Current-limiting fuse of the device blown.
Replace the ventilator and call for the
maintenance technician.
Содержание Puritan Bennett 540
Страница 1: ...Puritan Bennett 540 Ventilator User s Manual...
Страница 6: ...This page intentionally blank...
Страница 8: ...This page intentionally blank...
Страница 10: ...This page intentionally blank...
Страница 80: ...Installation and Assembly 4 20 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 86: ...Operating Procedures 5 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 94: ...This page intentionally blank...
Страница 100: ...Routine Maintenance 8 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Страница 124: ...Unpacking and Preparation C 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 140: ......