Specifications
A-6
Puritan Bennett 540™ Ventilator User’s Manual
Mode
Range: VOL A/C, PRES A/C, V SIMV, P SIMV, PSV/CPAP
Resolution: N/A
Accuracy: N/A
Default value: PRES A/C
Flow Pattern (flow pattern)
Range: Square, descending ramp
Resolution: N/A
cmH
2
O
Default value: Square, descending ramp
In V SIMV, flow pattern is set to square and is not adjustable
PEEP
Range: OFF (0.5 cmH
2
O) to 20 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy: ± (1 cm H
2
O + 10%) cmH
2
O
Default value: OFF
Rise time
Range: 1-4
Resolution: 1
Default value: 2
Depends on: Insp time
In P SIMV and V SIMV, rise time is set to 2 and is not adjustable.
Backup rate
Range: 5-40 bpm
Resolution: 1 bpm
Default value: 13
Depends on: Min Insp. time
In P SIMV and V SIMV, Backup rate = Max (8, R-Rate)
Apnea time
Range: AUTO or 1-60 s
Resolution: 1 s
Default value: AUTO
Depends on: Backup R
In PSV, Apnea time: AUTO = 60/Backup R s, or
Apnea time: AUTO = 30 s
In V SIMV or P SIMV, Apnea Time: AUTO = 12
Alarm settings
(based on breath mode)
Pressure A/C Mode
Low Inspired Tidal Volume
(Low VTI)
Range: 30 mL to 1990mL
Resolution: 10 mL
Default value: 300
Depends on: High VTI
High Inspired Tidal Volume
(High VTI)
Range: 80 mL to 3000 mL
Resolution: 10 mL
Default value: 2000 mL
Depends on: Low VTI
Pressure A/C Mode (Cont’d)
Table A-11.
Ventilator Range, Resolution, and Accuracy (Continued)
Setting or Monitored Parameter
Range, Resolution, and Accuracy
Содержание Puritan Bennett 540
Страница 1: ...Puritan Bennett 540 Ventilator User s Manual...
Страница 6: ...This page intentionally blank...
Страница 8: ...This page intentionally blank...
Страница 10: ...This page intentionally blank...
Страница 80: ...Installation and Assembly 4 20 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 86: ...Operating Procedures 5 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 94: ...This page intentionally blank...
Страница 100: ...Routine Maintenance 8 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Страница 124: ...Unpacking and Preparation C 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 140: ......