Ventilator Range, Resolution, and Accuracy
Setting or Monitored Parameter
Range, Resolution, and Accuracy
Ventilator Settings
Tidal volume (Vt)
Range: 50 mL to 2000 mL
Resolution: 10 mL
Accuracy: ± (10 ml + 10%) of setting
Default value: 500 mL
Depends on: Insp time, R-Rate
Pressure control
(P Control)
Range: 5 cmH
2
O to 55 cmH
2
0
Resolution: 1 cmH
2
O
Accuracy: ± (1 cmH
2
O + 10%) of P C PEEP setting
Default value: 15 cmH
2
O
Pressure support
(P Support)
Range: OFF or 5 cmH
2
O to 55 cmH
2
0
Resolution: 1 cmH
2
O
Accuracy: ± (1 cmH
2
O + 10%) of P S PEEP setting
Default value: 15 cmH
2
O
Inspiratory time (Insp Time)
Range: 0.3 s to 6.0 s for VOL A/C and PRES A/C
0.3 s to 4.0 s for V SIMV and P SIMV
Resolution: 0.1 s
Accuracy: ± 10%
Default value: 1.5 s
Depends on: R-Rate, Vt
Respiratory rate (R-Rate)
Range: 5 bpm to 60 bpm in VOL A/C and PRES A/C modes
1 bpm to 40 bpm in P SIMV and V SIMV modes
Resolution: 1 bpm
Accuracy: ± 1 bpm
Default value: 13
Depends on: High Rtot, Insp Time, Vt
Inspiratory sensitivity (Insp Sens)
Range: 1P-5
Resolution: 1
Accuracy: NA
Default value: 2
Exhalation sensitivity (Exh Sens)
Range: 5% to 95% of peak flow
Resolution: 5%
Accuracy: +/- (4 lpm +10%) of target exhalation flow based on
Exh Sense
Default value: 25%
In CPAP, V SIMV, and P SIMV, Exh Sens is fixed at 25% and is not
adjustable.
Ventilator Settings (Cont’d)
Содержание Puritan Bennett 540
Страница 1: ...Puritan Bennett 540 Ventilator User s Manual...
Страница 6: ...This page intentionally blank...
Страница 8: ...This page intentionally blank...
Страница 10: ...This page intentionally blank...
Страница 80: ...Installation and Assembly 4 20 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 86: ...Operating Procedures 5 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 94: ...This page intentionally blank...
Страница 100: ...Routine Maintenance 8 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Страница 124: ...Unpacking and Preparation C 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 140: ......