Installation and Assembly
4-8
Puritan Bennett 540™ Ventilator User’s Manual
Figure 4-7.
Single Limb Patient Circuit
9. To protect the exhalation port (as it will not be used in this configuration), place the cap (if
provided with the breathing circuit) over the exhalation port opening (item 10).
To connect a double limb circuit, do the following and refer to
1. Inspect the components of the patient circuit for any signs of damage, such as cracks (which might
cause leakage). Do not use damaged components to assemble the patient circuit.
2. Install the bacteria filter (item 1) on the TO PATIENT outlet port.
3. Attach one end of the short circuit tubing (item 4) to the filter (item 1).
4. Attach the other end of the circuit tubing to the humidifier’s inlet port (item 2).
5. Place a water trap (item 3) between the outlet port of the humidifier and the patient wye (item 5)
on the double limb circuit.
6. Place a water trap (item 3) between the patient wye and the inlet port of exhalation filter (item 5).
7. Connect one end of the small proximal pressure tubing (item 6) as close as possible to the patient
at the level of the double limb patient wye circuit connection (item 5) and the other end on the
ventilator patient pressure port.
8. Ensure that the water trap (item 3) on the expiratory limb of the patient circuit (item 7) is securely
installed between the patient wye (item 5) and the exhalation valve (item 9).
Note:
When shipped, the proximal pressure tube may already be connected to the patient wye. If so, simply verify that
the connection is secure and the tube shows no signs of damage, kinks, or obstructions.
LUNG
NOTE: Although shown here, the
humidifier (item 2), water trap
(item 3), and tubes upstream of
the single limb patient circuit are
not included with the ventilator.
Contact your supplier for more
information.
9
8
10
1
2
3
4
7
6
5
Содержание Puritan Bennett 540
Страница 1: ...Puritan Bennett 540 Ventilator User s Manual...
Страница 6: ...This page intentionally blank...
Страница 8: ...This page intentionally blank...
Страница 10: ...This page intentionally blank...
Страница 80: ...Installation and Assembly 4 20 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 86: ...Operating Procedures 5 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 94: ...This page intentionally blank...
Страница 100: ...Routine Maintenance 8 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Страница 124: ...Unpacking and Preparation C 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 140: ......