Range, Resolution, and Accuracy
Puritan Bennett 540™ Ventilator User’s Manual
A-11
High Minute Volume
(High Min Vol)
Range: 2 L to 99 L
Resolution: 0.5 L
Default value: OFF
Depends on: Vt, R-Rate, Low Min vol
High Respiratory Rate
(High Rtot)
Range: 10 bpm to 70 bpm
Resolution: 1 bpm
Default value: OFF
Depends on: R-Rate
Monitored Patient Data
Peak inspiratory pressure (PIP)
Range: 0 cmH
2
O to 99 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy: ± (0.2 cmH
2
O+ 8%)
Positive end expiratory pressure (PEEP)
Range: 0 cmH
2
O to 99 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy: ± (0.2 cmH
2
O + 8%)
Inspiratory tidal volume (VTI)
Range: 0 mL to 9,995 mL
Resolution: 5 mL
Accuracy: ± (10ml + 10%)
Expiratory tidal volume (VTE)
Range: 50 mL to 9,995 mL
Resolution: 5 mL
Accuracy: ± (10ml + 10%) for volumes > 50ml
A “—” symbol is displayed for volumes below 50ml
Inspiratory minute volume (Min VI)
Range: 0 L to 99.9 L
Resolution: 0.1 L
Accuracy: ± 20%
Expiratory minute volume (Min VE)
Range: 0 L to 99.9 L
Resolution: 0.1 L
Accuracy: ± 20%
Total respiratory rate (R tot)
Range: 0 bpm to 99 bpm
Resolution: 1 bpm
Accuracy: ± 1 bpm
Mean airway pressure (MAP)
Range: 0 cmH
2
O to 99 cmH
2
O
Resolution: 1
Accuracy: +/- (0.2 cmH2O + 8%)
I:E ratio (I:E)
Range: 9.9:1 to 1:9.9
Resolution: 0.1
Accuracy: ± (50ms or 10% whichever is greater)
Inspiratory time (I time)
Range: 0 s to 59.9 s
Resolution: 0.1 s
Accuracy: ± 50 ms or 5%, whichever is greater
Expiratory time (E time)
Range: 0 s to 59.9 s
Resolution: 0.1 s
Accuracy: ± 50 ms or 10%, whichever is greater
Table A-11.
Ventilator Range, Resolution, and Accuracy (Continued)
Setting or Monitored Parameter
Range, Resolution, and Accuracy
Содержание Puritan Bennett 540
Страница 1: ...Puritan Bennett 540 Ventilator User s Manual...
Страница 6: ...This page intentionally blank...
Страница 8: ...This page intentionally blank...
Страница 10: ...This page intentionally blank...
Страница 80: ...Installation and Assembly 4 20 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 86: ...Operating Procedures 5 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 94: ...This page intentionally blank...
Страница 100: ...Routine Maintenance 8 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Страница 124: ...Unpacking and Preparation C 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 140: ......