PN 4237523A
B-1
B
B
DIFF COMPARISON
B.1
INTRODUCTION
You can perform manual differentials as a measure of good QC practice or as recommended
by your laboratory, state, and federal protocols.
At the end of this appendix are blank log sheets used in the procedure.
B.2
PROCEDURE
1.
Collect two different normal blood samples. Follow NCCLS H3-A2 procedures.
The samples must be:
r
Collected in K
3
EDTA with 1.5 mg/mL of anticoagulant to blood
r
Maintained at room temperature
r
Within the reference range of:
NE 40 to 72%
LY 17 to 45%
MO 4 to 12%
EO 0 to 10%
BA 0 to1%
r
Analyzed between 30 minutes and 5 hours after draw.
2.
Label the samples.
Draw
1
2
3
4
5
6
7
8
9
10
11
12
1
2
Содержание HmX Hematology Analyzer
Страница 4: ...PN 4237523A iv REVISION STATUS ...
Страница 14: ...PN 4237523A xiv CONTENTS ...
Страница 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Страница 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Страница 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Страница 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Страница 77: ...PN 4237523A 6 3 REPORTING OPTIONS GRAPHIC FORMAT 6 Figure 6 3 Graphic Narrow Small Page Format ...
Страница 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Страница 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Страница 98: ......
Страница 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Страница 124: ...PN 4237523A 14 INDEX INDEX ...
Страница 126: ......