PN 4237523A
1-5
USE AND FUNCTION
METHOD HISTORY
1
Differential Measurement
The COULTER VCS established WBC differential technology using three measurements:
individual cell volume, high-frequency conductivity, and laser-light scatter.
The combination of low-frequency current, high-frequency current, and light-scattering
technology provides abundant cell-by-cell information that is translated by the instrument
into conventional stained film leukocyte categories. Correlation between the frequency of the
different cell types using stained film microscopy and this system is greater than 0.9 for
lymphocytes and granulocytes, and 0.7 for mononuclear cells.
Volume Analysis
Electronic leukocyte volume analysis, using low-frequency current, has been used since
1967.
18
It has been evaluated as a possible adjunct to the differential white cell
count.
19, 20, 21, 22
Conductivity Analysis
Cell walls act as conductors to high-frequency current. As the current passes through the cell
walls and through each cell interior, it detects differences in the insulating properties of cell
components. The current characterizes the nuclear and granular constituents and the
chemical composition of the cell interior.
23,
24,
25
Light Scatter Analysis
Beckman Coulter's experience in flow cytometry dates back decades to Fulwyler's pioneering
use of light scatter for cell analysis.
26
Loken et al. and Jovin et al. discuss the relationship of
particle size and refractivity to the angle of light scattered from a laser beam.
27
Reticulocyte (Retic) Analysis
Reticulocytes are immature, nonnucleated erythrocytes retaining a small network of
basophilic organelles, comprised of RNA and protoporphyrin. The enumeration of
reticulocytes provides a simple, effective means to determine red cell production and
regeneration.
28,
29,
30,
31
The most common means of measuring reticulocytes is to use supravital dyes, such as New
Methylene Blue or Brilliant Cresyl Blue. These dyes precipitate and aggregate the basophilic
substances within the reticulocyte, resulting in a granular, staining pattern easily seen with
light microscopy.
32
%
Analysis
Dennis B. Dorsey, MD, proposed in 1963 that the relatively constant blood cell indices could
be used to follow the performance of hematology instrumentation.
33
Brian Bull, MD,
improved the technique and named it
%
Analysis.
34
%
Analysis uses a "weighted moving average" of patient sample results because Koepke said
that QC materials "ideally should be similar in structure and in reactivity to the patient
constituent being measured, [and] therefore freshly drawn patient blood samples seem to be
the most appropriate [QC material]."
35
Bull explains, "The analyser [sic] is considered to be
‘in control’ when the MCV, MCH, and MCHC determined on a batch of 20 patients by use of
the
%
algorithm are within 3% of the expected mean indices of the population."
36
Содержание HmX Hematology Analyzer
Страница 4: ...PN 4237523A iv REVISION STATUS ...
Страница 14: ...PN 4237523A xiv CONTENTS ...
Страница 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Страница 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Страница 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Страница 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Страница 77: ...PN 4237523A 6 3 REPORTING OPTIONS GRAPHIC FORMAT 6 Figure 6 3 Graphic Narrow Small Page Format ...
Страница 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Страница 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Страница 98: ......
Страница 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Страница 124: ...PN 4237523A 14 INDEX INDEX ...
Страница 126: ......