PN 4237523A
3-6
OPERATION PRINCIPLES
SAMPLE FLOW
2.
The baths drain into the waste chamber (Figure 3.6).
Figure 3.6 Baths Draining
3.
Figure 3.7 applies to steps 3 through 7. The BSV rotates to segment three portions of the
sample:
r
1.6 µL for the RBC bath dilution
r
28 µL for the WBC bath dilution
r
28 µL for the WBC diff.
4.
The air pump transfers the diff portion to the sample line.
5.
Dispensers send diluent to the BSV to pick up and dilute the RBC and WBC portions.
6.
The CBC lytic reagent pumps send the CBC lytic reagent (LYSE S III or LYSE S 4) to the
WBC bath. It lyses the RBCs and converts the hemoglobin into a stable compound.
7.
Mixing bubbles enter both baths.
Hgb
lamp
Hgb
preamp
card
WBC
bath
Waste
chamber
RBC
bath
Содержание HmX Hematology Analyzer
Страница 4: ...PN 4237523A iv REVISION STATUS ...
Страница 14: ...PN 4237523A xiv CONTENTS ...
Страница 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Страница 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Страница 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Страница 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Страница 77: ...PN 4237523A 6 3 REPORTING OPTIONS GRAPHIC FORMAT 6 Figure 6 3 Graphic Narrow Small Page Format ...
Страница 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Страница 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Страница 98: ......
Страница 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Страница 124: ...PN 4237523A 14 INDEX INDEX ...
Страница 126: ......