PN 4237523A
7-4
BAR-CODE SPECIFICATIONS
BAR-CODE READER
Code-Related Specifications
See Table 7.2 for code-related specifications.
*According to American National Standard for bar-code specifications that yield 10 characters
per inch at NE = 0.0065".
Printing Methods
Photographic, thermal transfer, dot matrix, and laser printer.
7.2
BAR-CODE READER
Description
The HmX Hematology Analyzer uses a visible laser-type reader containing a Class II laser,
operating at a wavelength of 670 nm, with a maximum power output of 1 mW.
Settings and Defaults
Table 7.3 lists the available settings for the bar-code reader and the default setting for each
feature. To change a setting call your Beckman Coulter Representative. The
default
setting
appears in
bold
typeface.
Table 7.2 Code-Related Specifications
Code
Interleaved 2-of-5
Codabar/NW7
Code 39
Code 128
NE width
0.010 - 0.020"
0.010 Scaling*
Factor = 1.538
0.010 - 0.020"
0.010 - 0.020"
WE/NE ratio
(2.2 to 3):1
N/A
(2.21 to 3):1
N/A
Intercharacter gap
No
0.010" Min.
≥
NE
N/A
Data digits
2 to 16
3 to 16
3 to 16 (3 to 8 with
HEMOGARD tubes
3 to 16
Table 7.3 Bar-Code Reader Settings and Defaults
Feature
Interleaved 2 of 5
Codabar/NW7
Code 39
Code 128
Code Type
Enabled
/Disabled
Enabled
/Disabled
Enabled
/Disabled
Enabled
/Disabled
Fixed Length
N/A
Enabled
/Disabled
Enabled
/Disabled
Enabled
/Disabled
Code Length No. 1
0, 2, 4, 6, 8, 10,
12
,
14, 16
3, 4, 5, 6, 7, 8, 9,
10
,
11, 12, 13, 14, 15, 16
3, 4, 5, 6, 7, 8, 9,
10
,
11, 12, 13, 14, 15, 16
3, 4, 5, 6, 7, 8, 9,
10
,
11, 12, 13, 14, 15, 16
Code Length No. 2
0
, 2, 4, 6, 8, 10, 12,
14, 16
N/A
N/A
N/A
Check Digit
Enabled
/Disabled
Enabled
/Disabled
Enabled
/Disabled
N/A
Check Digit Output
Enabled
/Disabled
Enabled
/Disabled
Enabled
/Disabled
N/A
Check Digit Type
N/A
AIM-16
/MOD-16
(Coulter)/NW7
N/A
N/A
Intercharacter Gap
N/A
Enabled
/Disabled
Enabled
/Disabled
N/A
Start/Stop Match
N/A
Enabled
/Disabled
N/A
N/A
Содержание HmX Hematology Analyzer
Страница 4: ...PN 4237523A iv REVISION STATUS ...
Страница 14: ...PN 4237523A xiv CONTENTS ...
Страница 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Страница 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Страница 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Страница 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Страница 77: ...PN 4237523A 6 3 REPORTING OPTIONS GRAPHIC FORMAT 6 Figure 6 3 Graphic Narrow Small Page Format ...
Страница 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Страница 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Страница 98: ......
Страница 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Страница 124: ...PN 4237523A 14 INDEX INDEX ...
Страница 126: ......