37
Problem
Cause
Solution
The droplets are not
centered in the well plate.
Wrong well plate
vendor selected.
In the Surface tab, select the corresponding well
plate vendor for the most accurate positioning in
all wells.
Automatic calibration
is off-center.
Change to manual calibration to adjust the X, Y, and
Z set point exactly as desired.
The error “Connection to
the printer timed out” in
Corning® DNA Studio.
Connection to the
printer has been
broken.
Make sure the USB cable is connected to the device and the computer.
Corning Matribot®
Bioprinter SD card
mode is enabled.
Disable printing from SD card mode in the utilities tab in the Corning Matribot Bioprinter’s
LCD interface.
The Bioprint is not showing
up in the menu of the LCD
interface.
SD card is not
properly inserted.
Verify that the SD card is properly inserted in the SD card reader.
Corning Matribot
Bioprinter SD card
mode is not enabled.
Enable printing from SD card mode in the utilities tab in the Corning Matribot Bioprinter’s
LCD interface.
Control knob is not
responding when pressed.
The knob is pushed
all the way in.
Pull the knob gently towards you.