Manuale di Istruzioni
Instruction Manual
Rev.1 EN
23/08/2018
19
WARNING:
Considering the weight of the equipment and the footprint (see chapter 2.8), check the
maximum allowed floor load before installation.
WARNING:
Make sure to use proper fixing bolts according to the type of floor of the installation
place. CONTROLS cannot be considered responsible in case of injures to the operator or
damages to the equipment if the above instructions are not duly followed.
WARNING:
If the equipment is brought from a cold environment into a heated room, condensation on
and in the unit can constitute a danger and lead to malfunctioning of the unit when
started. Wait to connect and operate the unit until it is at room temperature.
WARNING:
The general grounding must comply with the rules in force; a wrong quality of the
grounding could be dangerous for the operator’s safety and cause bad function of the
electrical devices.
WARNING:
Due to the high pressure inside the hydraulic circuit (up to 650 bars) only tubes approved
by CONTROLS can be used. CONTROLS will not be responsible for damages to the
equipment or personnel injures in case of use of improper material and/or operations not
carried out by authorized technical personnel.
WARNING:
Covers/doors can only be removed/opened by maintenance people, by using relevant
tools/buttons. These tools and buttons are for use by maintenance personnel only. Never
leave the tools and buttons attached to the unit as this may endanger operator safety.
After performing maintenance/repair, make sure that all covers/doors are properly closed
and locked.
WARNING:
If the
TEST PARAMETERS
value is not properly set, it will cause improper functioning
of the machine (e.g. if the value is too high the machine will not be able to sense the
failure of the sample and therefore will continue pressing the sample and never stop).
This may result in damages to the equipment and/or danger for the operator. CONTROLS
will not be considered responsible in these events.
Содержание WIZARD 2 50-C13A0X
Страница 2: ......
Страница 37: ...Manuale di Istruzioni Instruction Manual Rev 1 EN 23 08 2018 35 Fig 2 9 Dimension and fixing of 50 C92A1X...
Страница 38: ...Manuale di Istruzione Instruction Manual 23 08 2018 Rev 1 EN 36 Fig 2 10 Dimension and fixing of 50 C92A2X...
Страница 118: ...Manuale di Istruzione Instruction Manual 23 08 2018 Rev 1 EN 116 Note...
Страница 119: ...Manuale di Istruzioni Instruction Manual Rev 1 EN 23 08 2018 117 Note...
Страница 120: ......