Manuale di Istruzioni
Instruction Manual
Rev.1 EN
23/08/2018
101
10.
Check the readings given by the
WIZARD 2
system against that of the reference
load cell. Normally this is done at values of 10%, 15%, 20%, 40%, 60%, 80% and
100% of the full scale of the loadcell. It is also possible to check the calibration at
lower loads if the compression frame has been so calibrated;
11.
If the percentage deviation is higher than permitted tolerances, the unit should be re-
calibrated by altering the calibration coefficient so as to obtain the best possible
linearity. To do this the edit calibration function should be accessed. After this the
verification process shall be repeated so as to check the new calibration.