Section 3: System Operation
996-0207
29
XXX Q-switch delay setting in microseconds. Red light labeled
Q-SW DELAY
µs
next to LED display should also be on. This value indicates the time between
the command to fire the flashlamp and the command to fire the Pockels cell.
The normal value is listed in the QA in front of the manual.
Pxx
Flashlamp discharge divider for laser output. Take the frequency of
flashlamp discharge and divide it by Pxx to get the frequency of laser output.
(Example: Fxx = F10 and Pxx = P05 then for every fifth discharge of the
flashlamps you get laser output. The system lases at 2 Hz) This feature allows
user to change the frequency of laser output without changing the flashlamp
discharge rate. The value of Pxx can be changed from 0-99 by pressing the
ARROW UP
or
ARROW DOWN
buttons. Normal value for Pxx is P01.
XXX Flashlamp discharge voltage in kilovolts. Red light labeled
VOLTS KV
next
to LED display should also be on. This indicates the charge voltage of the
laser head flashlamp capacitor. The normal value is listed in the QA in front
of the manual.
SoF
Indicates that the serial mode of operation through the RS232 port is off and
the system is running on its own internal clock commands.
Son
Indicates that the serial mode of operation through the RS232 port is on and
the system will await indicated input commands to operate. (See page
42
.)
EoF
Indicates that the external mode of operation through the
EXTERNAL
connector is off and the system is running on its own internal clock
commands.
Eon
Indicates that the external mode of operation through the
EXTERNAL
connector is on and the system will await indicated input commands to
operate. (See page
42
.)
Error messages
There are nine programmed error messages. When they appear, they are accompanied by
flashing and beeping. The definitions for the error messages are given below. Refer to
Table 12 Electrical troubleshooting
on page 71 to resolve these problems.
E00
Normal return.
E01
Surelite not in serial mode. Serial port is turned off. To operate in serial
mode, press
SELECT
then
ARROW UP
button.
E02
Coolant flow interrupted. Check water level and kinks in water lines.
E03
Coolant temperature over temp. Cooling water temperature is too high.
E04
(not used)
E05
Laser head problems.
E06
External interlock. External interlock has opened.
E07
End of charge not detected before lamp fire.
Содержание Surelite EX 10
Страница 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Страница 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Страница 26: ...Continuum Surelite Laser 16 996 0207...
Страница 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Страница 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Страница 74: ...Continuum Surelite Laser 64 996 0207 Figure 45 Cooling group chassis assembly for Surelite I 20 Hz...
Страница 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Страница 92: ...Continuum Surelite Laser 82 996 0207...
Страница 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Страница 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Страница 107: ...Appendix B Operator Notes...
Страница 108: ...Continuum Surelite Laser 98 996 0207...
Страница 109: ...Appendix B Operator Notes 996 0207 99...
Страница 110: ...Continuum Surelite Laser 100 996 0207...
Страница 111: ...Appendix B Operator Notes 996 0207 101...
Страница 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...