(1 ea.) Single Shot Cable that plugs into front of Power Cooling Unit
(PCU)
•
(1 ea.) 250V~ 16A power cord.
2. Tilt the laser bench and use a ruler to check the length of the 4 legs that
support the laser. All should be of equal length. If not, adjust the leg lengths to
ensure that the bench sits squarely on the table and does not rock.
3. Place the laser bench at the desired position on the table top. Use the “L”
brackets, foot pads,
¼
-20 bolts and washers from the accessory box to secure
the system to the table.
4. Locate the two water lines that run through the umbilical from laser. These
lines have standard garden hose male/female connectors.
•
At the rear of PCU locate the two hose fittings labeled
DI IN
and
DI
OUT
.
Connect the umbilical garden hose male-to-female connectors, to
the PCU panel garden hose connectors.
5. Take the 9 pin D connector from umbilical and plug it into the
LASER
HEAD
connector on PCU rear panel.
•
Locate ground strap from umbilical and connect to ground lug at rear of
PCU.
6. Take the head plug cable marked
FLASHLAMP
from rear of PCU, uncoil it
and plug into socket at rear of laser bench. Remember to tighten the twist
collar on the cable connector.
7. Determine the cooling group (CG) side of PCU. It will have a label on it that
reads “Open this side for cooling group maintenance”. Remove the 8 screws
holding CG access panel in place.
8. Lift panel up and off unit and put aside. Locate reservoir and fill to within 2
inches of top with distilled water.
9. Locate the power cord in the accessory box and install the plug to fit the 200-
240 Vac wall socket. The power cord consists of 3 wires, brown, blue and
green/yellow. Connect the wires as follows:
•
brown wire to a
HOT
•
blue wire to a
COMMON or HOT
•
green/yellow wire to
EARTH GROUND
.
•
The ac line voltage should be 200-240 V~. Phase relation between the 2
hots is not critical and can be between 120-180°.
•
Between the brown and blue wire the voltage measurement should be 200-
240 V~.
10.
Plug the square molded female end of ac line into 220 V~ plug on rear panel of
PCU marked “
220V~10A 50-60Hz” INPUT.
Содержание Surelite EX 10
Страница 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Страница 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Страница 26: ...Continuum Surelite Laser 16 996 0207...
Страница 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Страница 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Страница 74: ...Continuum Surelite Laser 64 996 0207 Figure 45 Cooling group chassis assembly for Surelite I 20 Hz...
Страница 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Страница 92: ...Continuum Surelite Laser 82 996 0207...
Страница 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Страница 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Страница 107: ...Appendix B Operator Notes...
Страница 108: ...Continuum Surelite Laser 98 996 0207...
Страница 109: ...Appendix B Operator Notes 996 0207 99...
Страница 110: ...Continuum Surelite Laser 100 996 0207...
Страница 111: ...Appendix B Operator Notes 996 0207 101...
Страница 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...