6. About Hardware
100
ADA16-32/2(PCI)F
About Calibration
Although this board is calibrated before shipping, you can use the calibration program to calibrate
analog input and output yourself.
Starting the calibration program
Click the [Calibration] button on the property page for the device to start the calibration program.
Proceed with connecting the calibration equipment and performing the calibration in accordance with
the instructions displayed by the calibration program.
Analog input calibration
Analog input calibration requires a reference voltage generator.
As the analog input has 16-bit resolution, use a reference voltage generator with a precision of at least 5
digits after the decimal point.
Calibrate one channel only for each range that you use.
Analog output calibration
Analog output calibration requires a digital multimeter.
As the analog output has 16-bit resolution, use a multimeter with a precision of at least 5 digits after the
decimal point.
Calibrate each channel separately for each range that you use.
Factory setting
You can use the calibration program to restore the factory calibration settings.
If for some reason you are unable to achieve the rated accuracy, please contact the CONTEC
information center.
* The name of the board
you have just added is
displayed.
- ADA16-32/2(PCI)F
Содержание ADA16-32/2(PCI)F
Страница 1: ...PC HELPER High Resolution Speed Analog I O Board for PCI ADA16 32 2 PCI F User s Guide CONTEC CO LTD ...
Страница 7: ...vi ADA16 32 2 PCI F ...
Страница 33: ...2 Setup 26 ADA16 32 2 PCI F ...
Страница 45: ...3 External Connection 38 ADA16 32 2 PCI F ...
Страница 99: ...5 About Software 92 ADA16 32 2 PCI F ...
Страница 108: ......