63
-
Prompt
Applicable
Patient Type
Occurring Condition
Alarm Level
ASYSTOLE
All patients
No QRS is detected for 4 consecutive seconds
User-selectable
VFIB /VTAC
Without
pacemaker
Fibrillatory wave for consecutive 4 seconds; or
The number of continuous Vent beats is larger than the
upper limit of cluster Vent beats (>5).
The RR interval is less than 600ms.
User-selectable
VT>2
Without
pacemaker
3 < the number of cluster PVCs < 5
User-selectable
COUPLET
Without
pacemaker
2 consecutive PVCs
User-selectable
BIGEMINY
Without
pacemaker
Vent Bigeminy
User-selectable
TRIGEMINY
Without
pacemaker
Vent Trigeminy
User-selectable
R ON T
Without
pacemaker
A type of single PVC under the condition that
HR<100,R-R interval is less than 1/3 the average
interval, followed by a compensating pause of 1.25X the
average R-R interval(the next R wave advances onto the
previous T wave).
User-selectable
PVC
Without
pacemaker
Single PVCs not belonging to the type of above
mentioned PVCs.
User-selectable
TACHY
All patients
5 consecutive QRS complex , RR interval is less than
500ms.
User-selectable
BRADY
All patients
5 consecutive QRS complex, RR interval is longer than
1.5s.
User-selectable
MISSED
BEATS
Without
pacemaker
When HR is less than 100 beats/min., no heart beat is
tested during the period 1.75 times of the average RR
interval; or
When HR is larger than 100 beats/min., no beat is tested
with 1 second.
User-selectable
PNP
With
pacemaker
No QRS complex and pacing pulse are availabe during
the period 1.75 times of the average R-R interval (only
considering patients with pacemaker.)
User-selectable
PNC
With
pacemaker
When pacing pulse is available, no QRS exists during
the period 1.75 times of the average RR interval (only
considering patients with pacemaker.)
User-selectable
Patient type:
All patients: refers to perform Arr.analysis on patients either with pacemakers or without pacemakers.
Without pacemaker: refers to perform Arr. Analysis only on the patients without pacemakers.
With pacemaker: refers to perform Arr. Analysis only on the patients with pacemakers.
Prompt message:
Message
Cause
Alarm
Level
ARR LEARNING
The QRS template building required for Arr. Analysis is in process.
No alarm
Note
Arrhythmia name displays in the Alarm Message Area.
11.9 Measuring RESP
11.9.1 How to measure RESP?
The monitor measures respiration from the amount of thoracic impedance between two ECG electrodes. The change of
impedance between the two electrodes, (due to the thoracic movement), produces a respiratory waveform on the
screen.
11.9.2 Setting Up RESP measurement
For RESP monitoring, it is not necessary for additional electrodes, however, the placing of electrodes is important.