116
-
Sensor inserted upside down.
Disconnect and reconnect he sensor with
the logos matching.
SpO
2
SENSOR
OFF
SpO
2
sensor may be disconnected
from the patient or the monitor.
Disconnect and reconnect the sensor.
Reattach sensor.
SpO
2
SENSOR FAULT
This message appears when the
sensor is faulty
Stop using the measuring function of SpO
2
module, notify biomedical engineer or our
service staff.
SpO
2
UNRECOGNIZED
SENSOR
board does not recognize the
sensor.
Make sure that the monitor and the patient
are in correct connection with the cables.
SpO
2
INCOMPATIBLE
SENSOR
This message is displayed when
the sensor is finding incompatible
sensor.
Make sure that the monitor use
incompatible sensor.
SpO
2
INTERFERENCE
Outside signal or energy
preventing reading.
Remove outside interference.
SpO
2
PULSE SEARCH
Unit is searching for the patients
pulse.
If values are not displayed within 30
seconds, disconnect and reconnect sensor.
If pulse search continues, remove sensor
and replace on a better perfused site.
SpO
2
LOW PERFUSTION
Signal too small.
Move sensor to better perfused site.
SpO
2
TOO
MUCH LIGHT
Too much light on patient(sensor).
Inadequate tissue covering sensor
detector.
Remove or reduce lighting. Cover sensor
from light.
Reposition sensor.
SpO
2
LOW
SIGNAL IQ
Low signal quality.
Ensure proper sensor application. Mover
sensor to a better perfused site.
SpO
2
BOARD FAULT
This message appears when the
Set board malfunctions.
Stop using the measuring function of SpO
2
module, notify biomedical engineer or our
service staff.
SpO
2
COMMUNICATION
ERROR
This message is displayed when
the front end module is having
problems communicating ( ie:
framing errors or bad checksums)
with the board.
Stop using the measuring function of SpO
2
module, notify biomedical engineer or our
service staff.
SpO
2
COMMUNICATION STOP
This message is displayed when
the host can not receive the data
from board for 5 seconds
Stop using the measuring function of SpO
2
module, notify biomedical engineer or our
service staff.
SpO
2
INIT ERR
This message is displayed when
the SpO
2
module initialization
error happened.
Stop using the measuring function of SpO
2
module, notify biomedical engineer or our
service staff.
"TEMP1 SENSOR OFF"
TEMP1 sensor is not connected
correctly.
Check the connection of TEMP1 sensor.
"TEMP2 SENSOR OFF"
TEMP2 sensor is not connected
correctly.
Check the connection of TEMP2 sensor.
"TEMP1 SENSOR OFF"
TEMP1 sensor is not connected
correctly.
Check the connection of TEMP1 sensor.
"TEMP2 SENSOR OFF"
TEMP2 sensor is not connected
correctly.
Check the connection of TEMP2 sensor.
"IBP1 LEAD OFF"
IBP1 sensor is not connected
correctly.
Check the connection of IBP1 sensor.
"IBP2 LEAD OFF"
IBP2 sensor is not connected
correctly.
Check the connection of IBP2 sensor.
"IBP1 NEED ZERO-CAL"
Zero calibrating must be done
Do zero calibrating for IBP1