Appendix II
System Alarm Prompt
PROMPT
CAUSE
MEASURE
"XX TOO HIGH"
XX value exceeds the higher
alarm limit.
Check if the alarm limits are appropriate and
the current situation of the patient.
"XX TOO LOW"
XX value is below the lower
alarm limit.
XX represents the value of parameter such as HR, ST1, ST2, RR, SpO
2
, IBP, NIBP, etc in the system.
"ECG WEAK SIGNAL"
The ECG signal of the patient is
too small so that the system can
not perform ECG analysis.
Check if the electrodes and lead wires are
connected correctly and the current
situation of the patient.
“NO PULSE”
The pulse signal of the patient is
too small so that the system can
not perform pulse analysis.
Check the connection of the sensor and the
current situation of the patient.
"RESP APNEA"
The respiration signal of the
patient is too small so that the
system cannot perform RESP
analysis.
Check the connection of the linking wire
and the current situation of the patient.
"CO
2
APNEA"
The respiration signal of the
patient is too small so that the
system cannot perform RESP
analysis.
Check the connection of CO
2
sensor and
the current situation of the patient.
"ASYSTOLE"
Patient suffers from Arr. Of
ASYSTOLE.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"VFIB/VTAC"
Patient suffers from Arr. of
VFIB/VTAC.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"COUPLET"
Patient suffers from Arr. of
COUPLET.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"BIGEMINY"
Patient suffers from Arr. Of
BIGEMINY.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"TRIGEMINY"
Patient suffers from Arr. of
TRIGEMINY.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"R ON T"
Patient suffers from Arr. of R
ON T.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"PVC"
Patient suffers from Arr. of PVC.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"TACHY"
Patient suffers from TACHY.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
" BRADY"
Patient suffers from BRADY.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"VT>2"
Patient suffers from Arr. of VT>2.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
“MISSED BEATS”
Patient suffers from Arr. of
MISSED BEATS.
Check the current situation of the patient.
Check the connection of the electrodes and
lead wires.
"PNP"
The pacemaker is not paced.
Check the connection of the pacemaker.