![Contec Medical Systems Co. CMS8000 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/contec-medical-systems-co/cms8000/cms8000_user-manual_2656215041.webp)
-
34
-
Chapter7 Recall
The monitor provides 480-hour trend data of all parameters, storage of 4800 NIBP measurement results and 71 alarm
events. This chapter gives detailed instruction for review of all data.
7.1 Trend Graph
The latest 1-hour trend is displayed every 1 or 5 seconds;
The latest 480-hour trend is displayed every 1, 5 or 10 minutes;
Pick "TREND GRAPH" in the SYSTEM MENU to call up the following menu:
Figure 7-1 Trend Graph Menu
The uppermost part is the name of the parameter, in which y-axis stands for value and x-axis time.
"
" Indicates the value of the parameter, which it points to, is below the x-axis, with corresponding time displayed
beyond the trend graph. Other trends except NIBP trend are displayed as continuous curves. three symbols "*" indicate
the position of the NIBP value ,including systolic value, diastolic value, mean value.
To select trend graph of a specific parameter:
Pick PARAM SELECT item (the first selection of the upper line) and select a requested parameter name by turning the
knob.
To select 1-hour or 480-hour trend graph:
Pick RESOLUTION item (the latter selection of the upper line), choose 1 or 5 sec for 1-hour trend graph and 1, 5 or 10
min for 480-hour trend graph.
To view other trend curves:
When "
" appears on the right part of the screen, pick "L-RIGHT" (the button at the extreme left of the lower line),
turn the knob clockwise to view later trend curves. When "
" appears on the left part of the screen, pick the same
item, turn the knob counterclockwise to view earlier trend curve.
To change the display scale
Pick the “ZOOM” button in the lower line to adjust the y-axis scale and thus change the trend curve in proportion. The
value beyond maximum value will be represented by the maximum value.
To obtain trend data of a specific time
The time to which the cursor points will change as the knob is turned. Parameter at this time is displayed below the
x-axis. When "
" appears on the right part of the screen, the trend graph pages down for later trend curve as the
cursor moves here. When "
" appears on the left part of the screen, the trend graph pages up for earlier trend curve as
the cursor moves here.
To print out the trend curve
Press REC button to print out the trend curve of current selected parameter.
Mark event
If an event is marked A, B, C, or D, then the corresponding event type will display on the axis time of the trend graph.
The event sign (A, B, C or D) is displayed in a frame.