18.2 Alarm and Measurement Default Settings
Settings are only entered once per table row if they are the same for all patient categories
1
.
18.2.1 Alarm Default Settings
Alarm Settings
Factory Default
ALARM VOL
1
ALM REC TIME
32 S
ALM PAUSE TIME
2 MIN
ALM TYPE
UNLATCH
KEYVOL
1
18.2.2 ECG, Arrhythmia, ST Default Settings
ECG Settings
Factory Defaults
Adult
Pedi
Neo
HR ALM
ON
ALM LEV
MED
ALM REC
OFF
ALM HI
120 bpm
160 bpm
200 bpm
ALM LO
50 bpm
75 bpm
100 bpm
HR FROM
AUTO
HR CHANNEL
CH1
LEAD TYPE
5 LEADS
SWEEP
25.0mm/s
Arrhythmia Settings
Factory Defaults
Adult
Pedi
Neo
ARR ANAL
OFF
PVCS ALM
OFF
ALM LEV
MED
ALM REC
OFF
ALM HI
10
Lead-independent ST Settings
Factory Defaults
Adult
Pedi
Neo
ST ANAL
OFF
ST ALM
OFF
ST ALM LEV
MED
ST ALM SEC
OFF
ST ALM HI
0.20
ST ALM LO
-0.20
18.2.3 Pulse Default Settings
Pulse Settings
Factory Defaults
Adult
Pedi
Neo
PR ALM
ON
PR ALM HI
120 bpm
160 bpm
200 bpm
PR ALM LO
50 bpm
75 bpm
100 bpm
SWEEP
25mm/s
AVG TIME
4S