![Contec Medical Systems Co. CMS8000 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/contec-medical-systems-co/cms8000/cms8000_user-manual_2656215062.webp)
-
55
-
Figure 11-6 ECG Setup Menu
ECG alarm setting
HR ALM: pick "ON" to enable prompt message and data record during the ECG alarm; pick "OFF" to disable the
alarm function, and there will be a
beside “ECG”.
ALM LEV: selectable from HIGH, MED,LOW. Level HIGH represents the most serious case.
ALM REC: pick "ON" to enable report printing upon ECG alarm.
ALM HI: used to set up the upper limit of ECG alarm.
ALM LO: used to set up the lower limit of ECG alarm.
ECG alarm is activated when the heart beat exceeds set ALM HI value or falls below ALM LO value.
HR alarm range:
Patient Type
MAX. ALM HI
MIN. ALM LO
Step
ADU
300
15
1
PED/NEO
350
15
1
Note
Please set the alarm limits according to clinical condition of individual patient. The upper limit shall not
exceed 20 beat/min higher than the patient's heart rate.
■
HR FROM
ECG, SpO
2
, AUTO and BOTH may detect heart rate. AUTO distinguishes heart rate source according to the quality of
signal. By picking ECG, the monitor prompts HR and activates HR beep. By picking SpO
2
, the monitor prompts
PULSE and activates pulse beep. BOTH mode displays HR and PR simultaneously, when this item is picked, PR
parameter is displayed to the right side of SpO
2
. As for the sound of HR or PR in BOTH mode, HR is given the priority,
i.e., if HR is available, whose sound will be sent out, but if HR is not available, then the sound will be for PR.
■
HR CHANNEL
"CH1"
to count the heart rate by CH 1 waveform
"CH2"
to count the heart rate by CH 2 waveform
"AUTO" the monitor selects a channel automatically
■
LEAD TYPE: used to select either 5 LEADS or 3 LEADS.
■
SWEEP
Available options for ECG SWEEP are 12.5, 25.0, and 50.0 mm/s.
■
ST ANALYSIS
Pick this item to access ST ANALYSIS menu, the detailed information about the menu is to be discussed in the
following section.
■
ARR ANALYSIS
Pick this item to access ARR ANALYSIS menu, the detailed information about the menu is to be discussed in the
following section.
■
OTHER SETUP
Pick this item to access ECG SETUP menu as shown below: