![Contec Medical Systems Co. CMS8000 Скачать руководство пользователя страница 118](http://html1.mh-extra.com/html/contec-medical-systems-co/cms8000/cms8000_user-manual_2656215118.webp)
-
111
-
80 to 120bpm < 8s
80 to 40bpm < 8s
Time to ALARM for tachycardia
Tachycardia ventricular:amplitude =1mV(p-v),heart rate = 206bpm
Ampitude
Average(s)
Range (s)
×1
2.7
2.2~ 2.9
×0.5
3.8
3.5~4.0
×2
3.8
3.5~4.0
Tachycardia ventricular :amplitude =2mV(p-v),heart rate = 195bpm
Ampitude
Average(s)
Range (s)
×1
2.3
2.0~3.0
×0.5
3.3
3.0~4.0
×2
2.2
2.0~3.0
Pace Pulse
Pulse indicator
Pulse is marked if the requirements of ANSI/AAMI
EC13:2002,Sect.4.1.4.1 are met:
Amplitude: ±2mV ~ ±700mV
Width: 0.1ms ~ 2ms
Pulse Rejection
Pulse is rejected if the requirements of ANSI/AAMI
EC13-2002:Sect.4.1.4.1 are met:
Amplitude: ±2mV ~ ±700mV
Width: 0.1ms ~ 2ms
2.8 RESPIRATION
Method
Impedance between R-F(RA-LL)
Differential Input Impedance
>2.5 MΩ
Measuring Impedance Range:
0.3~5.0Ω
Base line Impedance Range:
0.1 KΩ– 2.5 KΩ
Bandwidth
0.3 ~ 2.5 Hz
Scan speed
6.25mm/s, 12.5mm/s, 25mm/s
Resp Rate
Measuring and Alarm Range
Adult
0 ~ 120 rpm
Neonate/Pediatric
0 ~ 150 rpm
Resolution
1 rpm
Accuracy
±2 rpm
Apnea Alarm
10 ~ 40 s
2.9 NIBP
Method
Oscillometric
Mode
Manual, Auto, STAT
Measuring Interval in AUTO Mode
1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120,240,480,960 Min
Measuring Period in STAT Mode
5 Min
Alarm Type
SYS, DIA, MEAN
Measuring and alarm Range
Adult Mode
SYS
40 ~ 270 mmHg
DIA
10 ~ 215 mmHg
MEAN
20 ~ 235 mmHg
Pediatric Mode
SYS
40 ~ 200 mmHg
DIA
10 ~ 150 mmHg
MEAN
20 ~ 165 mmHg
Neonatal Mode
SYS
40 ~ 135 mmHg
DIA
10 ~ 100 mmHg
MEAN
20 ~ 110 mmHg
Resolution
Pressure
1mmHg
Accuracy