NOTE:
If the following measures fail to alleviate the problem, please call
the distributors.
Troubles
Possible causes
Solutions
1. No response
after start the
device.
1. Lack of power;
2. Fuse might be
burned out.
1. Check whether the power
plug and socket of the
instrument are plugged or
not;
2. Replace the burned fuse
with
a
new
one
(Specifications: F2AL250V)
2.
The
UE
pedals
are
unable to rotate.
1. You had select the
Leg training;
2. Opposite to the set
direction.
1. Select the Arm training;
2.
Rotate
in
the
set
direction.
3.
The
LE
pedals
are
unable to rotate.
1. You had select the
Arm training;
2. Opposite to the set
direction.
1. Select the Leg training;
2.
Rotate
in
the
set
direction.
4.
Excessive
noise
during
operation.
1. The position of the
trainer is not stable;
2. The big screw
knob is not tightened.
1. Place the trainer in a
smooth position for training;
2. Tighten the big screw
knob.