Guidance and manufacture’s declaration – electromagnetic
immunity
The DQAPT
is intended for use in the electromagnetic environment
specified below. The customer or the user of DQAPT
should assure
that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
±8 kV contact
±2,4,8,15 kV
air
±8 kV contact
±2,4,8,15 kV
air
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical
fast
transient/
burst
IEC
61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
Mains
power
quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC
61000-4-5
± 0.5kV, ± 1
kV line(s) to
lines
± 0.5kV, ± 1
kV, ± 2 kV
line(s) to
earth
± 0.5kV, ± 1
kV line(s) to
lines
± 0.5kV, ± 1
kV, ± 2 kV
line(s) to
earth
Mains
power
quality
should be that of a
typical commercial or
hospital environment.
Voltage
dips, short
interruption
0% U
T
; 0.5
cycle (0°,45°,
90°,135°,180
0% U
T
; 0.5
cycle (0°,45°,
90°,135°,180
Mains
power
quality
should be that of a
typical commercial or