- 23 -
Example:
Set value programmed at the circulator
T
set
= 70.7°C
Actual temperature in bath / system
T
act
= 70.5°C
Deviation, calculated according to
ΔT = T
set
− T
act
ΔT = 0.2°C
Entry of the corrected value ΔT as
correction factor “cS”
cS = +0.2°C
The temperature control is thus internally altered so that the
desired 70.7°C is also attained in the external system. The
temperature displayed at the circulator and that of the external
system now correspond with each other.